Sanyo Welder MCO 19MUVH User Guide

INSTRUCTION MANUAL  
Multi Gas Incubator  
MCO-19M(UVH)  
MCO-19M(UV)  
MCO-19M  
 
CONTENTS  
SPECIFICATIONS  
P. 74  
P. 75  
P. 76  
PERFORMANCE  
SAFETY CHECK SHEET  
2
 
INTRODUCTION  
Read this manual carefully before using the Product and follow the instructions for safety operation.  
Sanyo disavows any responsibility for safety if the Product is used for other than the intended use or  
used with any procedures other than those given in this manual.  
Keep this manual in a suitable place so that it can be referred to as necessary.  
The contents of this manual are subject to change without notice for improvement of performance or  
functions.  
Contact a Sanyo sales representative or agent if any page of the manual is lost or the page order is  
incorrect.  
Contact a Sanyo sales representative or agent if any point in this manual is unclear or if there are any  
inaccuracies.  
No part of this manual may be reproduced in any form without the expressed written permission of  
Sanyo.  
CAUTION  
SANYO guarantees the product under certain warranty conditions. SANYO in no way shall be  
responsible for any loss of content or damage of content.  
3
 
PRECAUTIONS FOR SAFE OPERATION  
It is imperative that the user complies with this manual as it contains  
important safety advice.  
Items and procedures are described so that you can use this unit correctly and safely.  
If the precautions advised are followed, this will prevent possible injury to the user and  
any other person.  
Precautions are illustrated in the following way:  
WARNING  
Failure to observe WARNING signs could result in a hazard to personnel  
possibly resulting in serious injury or death.  
CAUTION  
Failure to observe CAUTION signs could result in injury to personnel and  
damage to the unit and associated property.  
Symbol shows;  
this symbol means caution.  
this symbol means an action is prohibited.  
this symbol means an instruction must be followed.  
Be sure to keep this manual in a place accessible to users of this unit.  
< Label on the unit >  
This mark is labeled on the cover in which the electrical components of high voltage  
are enclosed to prevent the electric shock.  
The cover should be removed by a qualified engineer or a service personnel only.  
WARNING  
As with any equipment that uses CO2 gas, there is a likelihood of oxygen depletion in the vicinity  
of the equipment. It is important that you assess the work site to ensure there is suitable and  
sufficient ventilation. If restricted ventilation is suspected, then other methods of ensuring a  
safe environment must be considered. These may include atmosphere monitoring and  
warning devices.  
4
 
WARNING  
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to  
rain water.  
Only qualified engineers or service personnel should install the unit. The installation by  
unqualified personnel may cause electric shock or fire.  
Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from  
turning over. If the floor is not strong enough or the installation site is not adequate, this may result  
in injury from the unit falling or tipping over.  
Never install the unit in a humid place or a place where it is likely to be splashed by water.  
Deterioration of the insulation may result which could cause current leakage or electric shock.  
Never install the unit in a flammable or volatile location. This may cause explosion or fire.  
Never install the unit where acid or corrosive gases are present as current leakage or electric  
shock may result due to corrosion.  
Always ground (earth) the unit to prevent electric shock. If the power supply outlet is not  
grounded, it will be necessary to install a ground by qualified engineers.  
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such  
grounding may cause electric shock in the case of an incomplete circuit.  
Connect the unit to a power source as indicated on the rating label attached to the unit. Use  
of any other voltage or frequency other than that on the rating label may cause fire or electric shock.  
Never store volatile or flammable substances in this unit if the container cannot be sealed. These  
may cause explosion or fire.  
Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet on the unit.  
This may cause electric shock or injury by accidental contact with moving parts.  
Use this unit in safe area when treating the poison, harmful or radiate articles. Improper use  
may cause bad effect on your health or environment.  
Turn off the power switch (if provided) and disconnect the power supply to the unit prior to any  
repair or maintenance of the unit in order to prevent electric shock or injury.  
Do not touch any electrical parts (such as power supply plug) or operate switches with a wet  
hand. This may cause electric shock.  
5
 
WARNING  
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of  
maintenance. These may be harmful to your health.  
Never splash water directly onto the unit as this may cause electric shock or short circuit.  
Never put containers with liquid on the unit as this may cause electric shock or short circuit when  
the liquid is spilled.  
Never bind, process, or step on the power supply cord, or never damage or break the power  
supply plug. A broken supply cord or plug may cause fire or electric shock.  
Do not use the supply cord if its plug is loose. Such supply cord may cause fire or electric shock.  
Never disassemble, repair, or modify the unit yourself. Any such work carried out by an  
unauthorized person may result in fire, or electric shock or injury due to a malfunction.  
Disconnect the power supply plug if there is something wrong with the unit. Continued  
abnormal operation may cause electric shock or fire.  
When removing the plug from the power supply outlet, grip the power supply plug, not the cord.  
Pulling the cord may result in electric shock or fire by short circuit.  
Disconnect the power supply plug before moving the unit. Take care not to damage the power  
cord. A damaged cord may cause electric shock or fire.  
Disconnect the power plug when the unit is not used for long periods. Keeping the connection  
may cause electric shock, current leakage, or fire due to the deterioration of insulation.  
If the unit is to be stored unused in an unsupervised area for an extended period, ensure that  
children do not have access and that doors cannot be closed completely.  
The disposal of the unit should be accomplished by appropriate personnel. Remove doors to  
prevent accidents such as suffocation.  
Do not put the packing plastic bag within reach of children as suffocation may result.  
Use the reagent specified by Sanyo for H2O2 decontamination. Using a different H2O2 solution  
may result in explosion or damage to the Incubator.  
When performing H2O2 decontamination, securely close gastight split doors, inner door and outer  
door. Failure to do so may be harmful to health due to leakage of H2O2 gas.  
During H2O2 decontamination, plug the access hole with the silicon cap that is provided. Failure  
to do so may be harmful to health due to leakage of H2O2 gas.  
6
 
CAUTION  
Use a dedicated power source (a dedicated circuit with a breaker) as indicated on the rating label  
attached to the unit. A branched circuit may cause fire resulting from abnormal heating.  
Connect the power supply plug to the power source firmly after removing the dust on the plug.  
A dusty plug or improper insertion may cause a heat or ignition.  
Never store corrosive substances such as acid or alkali in this unit if the container cannot be  
sealed. These may cause corrosion of inner components or electric parts.  
Check the setting when starting up of operation after power failure or turning off of power  
switch. The stored items may be damaged due to the change of setting.  
Be careful not to tip over the unit during movement to prevent damage or injury.  
Prepare a safety check sheet when you request any repair or maintenance for the safety of service  
personnel.  
Wear rubber gloves when handling the H2O2 reagent. Direct contact with the H2O2 reagent may  
result in inflammation of the skin.  
H2O2 decontamination can be performed only for the chamber and chamber attachments with  
standard specifications, and not for any other objects.  
Perform H2O2 decontamination with the chamber attachments arranged as specified by Sanyo.  
Arranging them in a different way may result in insufficient decontamination.  
After H2O2 decontamination has been completed, wear rubber gloves and use a non-woven cloth  
to wipe off the residual H2O2 fluid from the bottom of the chamber, any objects that were  
decontaminated, and the bottoms of ducts.  
7
 
LABELS ON THE INCUBATOR  
Warning and caution labels are attached to the Incubator. The following table describes the labels.  
This label is attached to covers that access high-voltage electrical components to  
prevent electric shock. Only a qualified engineer or service personnel should be  
allowed to open these covers.  
This symbol indicates an ultraviolet light (UV) caution.  
This symbol indicates that caution is required. Refer to product documentation for  
details.  
This symbol indicates a hot surface.  
This symbol indicates an earth.  
This symbol means “ON” for a power switch.  
This symbol means “OFF” for a power switch.  
ENVIRONMENTAL CONDITIONS  
This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1):  
Indoor use;  
Altitude up to 2000 m;  
Ambient temperature 5oC to 40oC  
Maximum relative humidity 80% for temperature up to 31oC decreasing linearly to 50% relative humidity  
at 40oC;  
Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage;  
Other supply voltage fluctuations as stated by the manufacturer;  
Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains  
supply the minimum and normal category is II;  
Pollution degree 2 in accordance with IEC 60664.  
8
 
INCUBATOR COMPONENTS  
Handles  
6
3
2
9
10(inside)  
13  
5
7
1
8
16, 17(inside)  
For MCO-19M(UVH)/19M(UV) or  
when MCO-19UVS is installed.  
Handles  
12 11  
25  
19  
24  
15  
14  
4
For MCO-19M(UVH)/19M(UV) or  
when MCO-19UVS is installed.  
21  
22  
23  
18  
20  
Rear Right Side  
Rear Left Side  
9
 
1. Outer door: The outer door is held to the frame with a magnetic seal. A door heater is installed in the  
door panel. The door opening is reversible. Contact a Sanyo representative or agent to change the door  
hinge from left to right or vice versa.  
2. Inner door: Open when take the trays out or clean inside.  
3. Gastight split door: Made of tempered glass, however avoid excessive impact on the glass.  
4. Leveling feet: The leveling feet can be turned to adjust the height. Adjust the feet so that the Incubator  
is level.  
5. Trays: Set trays at the same level as each gastight split door. Open a inner door when take the  
tray out.  
6. Tray supports: The tray supports can be removed by lifting the front side and pulling toward you.  
7. Side supports: The right and left side supports can be removed for disinfection. Refer to page 45.  
8. Duct: The duct for the path for circulating air. It is removable.  
9. Fan cover: The fan cover serves as the inlet for circulating air. It is removable.  
10. Fan (inside the duct): The fan is made from polypropylene resin. It can be disinfected in an  
autoclave.  
11. Sample air outlet: The sample air outlet also functions as an internal gas outlet. Normally, cover this  
outlet with the sample air outlet cap.  
12. Sample air outlet cap: Always attach this cap except when using the sample air outlet.  
13. Door switch: Detects the door opening/closing and stops the circulating fan and electromagnetic  
valve for CO2 and N2/O2 when door is open. UV lamp is also deactivated by door opening*.  
14. Humidifying pan: Fill the humidifying pan with sterile distilled water. Install the pan properly so that it  
can be covered with the pan cover.  
15. Humidifying pan cover: The cover prevents UV light entering the chamber. When filling the pan, lift  
the front side and take out the pan. Refer to page 48 for details.  
16. UV lamp*: This Sanyo UV lamp does not generate ozone. Never look directly at the UV light. For  
replacement, contact a Sanyo representative or agent.  
17. Water level sensor for humidifying pan: This sensor detects the water level in the humidifying pan.  
Refer to page 44 for details.  
18. Remote alarm terminals: Refer to page 13.  
19. Access port: Place the silicon caps on the port both outside and inside when the port is not being  
used.  
20. Power switch: This is the main switch for the Incubator. It also functions as an overcurrent breaker.  
21. Glow starter*: The glow started if for the UV lamp (model FG-7P)  
22. Connecting port A/B for CO2 gas pipe: When the optional MCO-21GC Automatic CO2 Cylinder  
Changeover System is installed, both ports A and B are available. If the MCO-21GC is not installed, only  
port A is available. Refer to page 16 for gas cylinder connection. Ensure that the gas pressure is set at  
0.03 MPaG (0.3 kgf/cm2G, 4.3 psiG). Refer to page 67 for automatic CO2 cylinder changeover.  
10  
 
23. Connecting port A/B for N2/O2 gas pipe: Refer to page 17 for gas cylinder connection. Ensure that  
the gas pressure is set at 0.05 MPaG (0.5 kgf/cm2G, 7.1 psiG). Refer to page 30 for automatic N2 (or O2)  
gas cylinder changeover.  
24. Attachment location for electric key: This is the attachment location for the electric key included in  
the optional Component H2O2 Decontamination Kit (MCO-HL). This kit must be attached to perform H2O2  
decontamination. Refer to the installation procedure for MCO-HL for details.  
25. Gas injection nozzle  
* MCO-19M(UVH)/19M(UV) or when an optional UV System Kit (MCO-19UVS) is installed.*  
11  
 
Control panel and keypad  
LCD panel  
2
1
10  
3
4
9
7
8
5
6
1. Upper limit regulator:  
6. CE Key:  
This regulator is used to set the upper  
temperature limit.  
Press this key to clear the entered value  
when entering a setting.  
2. OVERHEAT indicator:  
7. Cursor Keys (Up, Down, Left, and Right):  
Use these keys to move the cursor on the  
LCD panel.  
This indicator lights when the chamber  
temperature reaches the upper limit.  
3. MENU Key:  
8. ENTER Key:  
Press this key to access the menu.  
Press this key to select a menu command.  
When entering a setting, pressing this key to  
move to the next parameter.  
4. Contrast knob:  
Turn this knob to adjust the contrast of the  
LCD.  
9. Numeric Keys  
5. BUZZER Key:  
10. H2O2 Key:  
Press this key to silence the buzzer.  
Note: It is not possible to silence the  
buzzer for an upper limit temperature  
alarm.  
Press this key to start H2O2 decontamination.  
Refer to page 49 for details.  
The following optional components must be installed to perform decontamination.  
UV Lamp Add-on Kit (MCO-19UVS) (This is provided as standard equipment for the MCO-19M(UV).)  
H2O2 Decontamination Kit (MCO-HL)  
H2O2 Generator (MCO-HP)  
12